OCTIV | VI Technology – Standards of Calibration

Octiv VI Probe Technology -- Voltage | Current | Phase | Impedance | Harmonics

INTRODUCTION

High power radio-frequency (RF) voltage and current sensors need to be accurately calibrated to a traceable standard. Calibrating to high accuracy can be the most challenging aspect of high power, voltage-current sensor manufacture. This is due to the many sources of error in any calibration process. If the calibration is performed accurately and correctly, then most errors can be characterized and removed.

IMPEDANS CALIBRATION PROCESS

The Octiv VI probe incorporates two separate calibration procedures, one for frequency and one for temperature calibration. Frequency calibration is the most complex and crucial of the two. It uses multiple measurements against traceable standards to calibrate the output of the Octiv sensor. The temperature calibration ensures that when no inputs change (except device temperature), the output remains constant.

Frequency calibration

Due to the mechanical and electrical structure of most RF sensors, the actual measurement ports are physically removed from the true measurement points by amplifiers, couplers, extension cables, filters, etc. and each one in turn adds an error to the measurement. Therefore, to provide high accuracy these RF sensors need to be calibrated at the reference ports. The process of calibration usually includes several measurements planes with different well known standards. These standards generally are traceable to National Institute of Standards and Technology (NIST). Commonly used standards are open, short and load (50 Ω). To perform calibration, the output port of the sensor (which is a reference plane) is periodically connected to each standard. The raw voltage and current values from the uncalibrated sensor are measured. Open and short measurements are comparably simple to Implement. The last and probably the most important calibration step is the load measurement where the power value from the power meter is used. This measurement is required to define the complex values of voltage and
current at the reference plane.


Equation (1)
Equation 1



Equation (2)
Equation 2


Vrms and Irms are rms values of voltage and current and PLoad and ZLoad are absorbed power and input impedance of the load at the reference measurement plane. From equation (1) and equation (2) for Vrms and Irms, it follows that knowing exactly the PLoad and ZLoad values are crucial. To measure PLoad it is most common for a high power thermal power meter to be used. These are usually limited to an accuracy close to 1 %. In regard to ZLoad, it is generally assumed constant at 50 Ω with a small error which is measured and usually expressed in the data-sheet as Voltage Standing Wave Ratio (VSWR). For calibrations at low powers, the dependence of ZLoad on absorbed power can be assumed negligible but for calibrations at high powers (from 100 W to several kW) conditions are changing. Due to the physical characteristics of any RF load or attenuator, the impedance and attenuation are functions of absorbed power. If the sensor is going to be calibrated at multiple frequencies and multiple power levels, the total load measurements can easily take more than 10 minutes. This shows that before each calibration step, the load must be brought to thermal equilibrium. A simplified error expression for this method is:


Equation (3)
Equation 3



Equation (4)
Equation 4


where ΔPLoad is the power meter measurement error and ΔZLoad is the input impedance error of the power meter. From these equations, it follows that significant uncertainty can arise, even when using a very precise reference power meter, if the calibration process is not designed correctly. To overcome this problem, Impedans is using a different calibration approach.


Figure1. Frequency calibration set-up.


The setup for this frequency calibration is shown in Figure 1. In normal operation, the Octiv sensor collects two independent measurements at the true measurement points related to voltage and current for each frequency and each power level. To transform these two measurements to actual voltage and current values the following equations are used:


Equation (5)
Equation 5



Equation (6)
Equation 6


where V and I are actual complex voltage and current at the reference plane. X and Y are complex values of digitised signals at the input of the analogue-to-digital converter. A, B, C and D are unknown complex parameters that need to be defined by calibrating the sensor at reference plane using three standards (open, short and 50 Ω) which all are traceable to NIST. Each equation has two unknowns and therefore multiple measurements for different load impedances are required to evaluate these unknowns. For each frequency and power level, the Octiv calibration uses load and open circuit measurements to evaluate parameters A and B. Similarly, the 50 Ω load and open circuit measurements to evaluate the C and D parameters. The boundary conditions of V and I for load measurement are defined from equation (1) and equation (2). The load measurements are obtained from the power meter which consists of a high power attenuator (gain uncertainty is periodically analysed with a vector network analyser (VNA), not shown in Figure 1) and a precision low power thermal power meter (uncertainty 0.04 dB (0.92 %), relative uncertainty 0.01 dB (0.23 %)). The measured power (P), voltage (V) and current (I) values can be expressed as:


Equation (7)
Equation 7



Equation (8)
Equation 8



Equation (9)
Equation 9



Equation (10)
Equation 10


where ΔGAtt is the attenuator gain error, ΔPMeter is the power meter error and ΔZAtt is the attenuator impedance error. On top of this, we must add the unit measurement repeatability error (ΔOE). Therefore, the accuracy of V and I becomes:


Equation (11)
Equation 11



Equation (12)
Equation 12



Figure 2. Graph of Attenuator Gain Uncertainty v’s Frequency.


The octiv unit repeatability is 0.1 %. As seen in Figure 2, the attenuator gain uncertainty is 0.5 %. The accuracy of the thermal power meter used is 0.23 % The uncertainty of the attenuator impedance is shown in Figure 3 to be 0.2 % below 60 MHz. This gives a total voltage and current uncertainty for the Octiv sensor of 1.03 %. Therefore, the total uncertainty for power measurement into a 50 Ω load is ± 1 %.


Figure 3. Attenuator Impedance Uncertainty v’s Frequency.


Phase error is simply due to the error of the attenuator impedance. Again the measurement repeatability of the octiv must be included:


Equation (13)
Equation 13


For phase measurements the octiv measurement repeatability is 0.1°. The phase uncertainty of the attenuator impedance as shown in Figure 4 is 0.25° below 100 MHz. This gives the Octiv a phase uncertainty of 0.35°.


Figure 4. Attenuator Impedance Phase Uncertainty v’s Frequency.


Temperature calibration

The setup for the temperature calibration is shown in Figure 5. A power source supplies a constant power through the sensor, to a 50 Ω load, in a temperature controlled environment As previously mentioned, when power is applied to the 50 Ω load, it is necessary to wait until the load has reached thermal equilibrium and load impedance is no longer changing in order to assume the power is constant.


Figure 5. Octiv temperature calibration set-up.


At this point, the output from the Octiv is monitored and the temperature in the temperature controlled environment is ramped up. The error in the output due to temperature can then be plotted and the temperature calibration coefficients are stored in the Octiv memory. This calibration reduces the Octiv voltage and current measurement errors to ± 0.2 % due to temperature.